Med Phys2021;48(9):4993-5003.September 1, 2021Journal Article

Technical Note: Scanner dependence of adaptive statistical iterative reconstruction with 3D noise power spectrum central frequency and noise magnitude ratios.

Akira Hasegawa1,2, Toshihiro Ishihara1, Matthew Allan Thomas3, Tinsu Pan3
1Department of Radiological Technology, National Cancer Center Japan, Tokyo, Japan.
2AlgoMedica, Inc., Sunnyvale, California, USA.
3Department of Imaging Physics, M.D. Anderson Cancer Center, University of Texas, Houston, Texas, USA.

Abstract

PURPOSE: In this study, the noise reduction properties of the adaptive statistical iterative reconstruction (IR) on two different CT scanners of 64 and 256-slice were compared and their differences were assessed. METHODS AND MATERIALS: The homogeneous module of the ACR CT phantom was scanned on the 64 and 256 slices CT scanners from the same vendor in the range of 15-40 mA. On each scanner, the d…

Create a free account to keep reading

Free members get 10 full research views every month across publications, clinical trials, FDA clearances, adverse events, and NIH grants. No credit card required.

Want unlimited research access? See Pro plans

Data Accuracy Notice: Research intelligence on Health AI Central is aggregated from public sources (PubMed, ClinicalTrials.gov, FDA, NIH, CMS, and others) and refreshed nightly. Classifications and derived metrics are produced by automated methods described in our Methodology. We recommend verifying critical data points against the primary sources before making decisions.